4.6 Article

Column selectivity in reversed-phase liquid chromatography - V. Higher metal content (type-A) alkyl-silica columns

期刊

JOURNAL OF CHROMATOGRAPHY A
卷 1026, 期 1-2, 页码 77-89

出版社

ELSEVIER
DOI: 10.1016/j.chroma.2003.11.043

关键词

column seletivity; stationary phases, LC; selectivity; alkyl-silica column; bonded-zirconia column

向作者/读者索取更多资源

Retention measurements involving 16 test solutes have been carried out for 38 type-A alkyl-silica columns and three bonded-zirconia columns. These measurements have been analyzed in terms of a model previously developed for type-B columns, so as to yield values of five column selectivity parameters (H, S*, A, B, C) for each type-A column. Overall differences in selectivity between type-A and -B columns can be related to the average values of H, S* etc. for each column type. Compared to type-B columns, type-A columns provide generally stronger retention for carboxylic acids, while solutes that are more hydrophobic or less bulky are more retained on type-B columns. Hydrogen-bond acceptors (e.g. aliphatic amides) and cations (e.g. protonated bases) are strongly retained on type-A versus type-B columns. Compared to type-B columns, bonded-zirconia columns show much increased retention of cations and reduced retention of hydrogen-bond acceptors. Because of relatively large differences in the selectivity of bonded-zirconia, type-A, and type-B columns, it will prove difficult to find columns of different type (e.g. a type-A and a type-B column) which have equivalent selectivity. Type-A columns also tend to be more different from each other (in terms of selectivity) than is the case for type-B columns. As a result, the replacement of a given type-A column by an equivalent type-A column also appears unlikely, except for samples that do not contain ionized compounds. (C) 2003 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据