Dielectrophoretic registration of living cells to a microelectrode array

期刊

BIOSENSORS & BIOELECTRONICS
卷 19, 期 7, 页码 771-780

出版社

ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/j.bios.2003.08.013

关键词

dielectrophoresis; endothelial cells; microfabrication; array; biosensor; patterning

向作者/读者索取更多资源

We present a novel microfabricated device to simultaneously and actively trap thousands of single mammalian cells in alignment with a planar microelectrode array. Thousands of 3 mum diameter trapping electrodes were fabricated within the bottom of a parallel-plate flow chamber. Cells were trapped on the electrodes and held against destabilizing fluid flows by dielectrophoretic forces generated in the device. In general, each electrode trapped only one cell. Adhesive regions were patterned onto the surface in alignment with the traps such that cells adhered to the array surface and remained in alignment with the electrodes. By driving the device with different voltages, we showed that trapped cells could be killed by stronger electric fields. However, with weaker fields, cells were not damaged during trapping, as indicated by the similar morphologies and proliferation rates of trapped cells versus controls. As a test of the device, we patterned similar to20,000 cells onto a 1 cm(2) grid of rectangular adhesive regions, with two electrodes and thus two cells per rectangle. Our method obtained 70 +/- 1% fidelity versus 17 +/- 1% when using an existing cell-registration technique. By allowing the placement of desired numbers of cells at specified locations, this approach addresses many needs to manipulate and register cells to the surfaces of biosensors and other devices with high precision and fidelity. (C) 2003 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据