4.7 Article Proceedings Paper

Propofol-imprinted membranes with potential applications in biosensors

期刊

ANALYTICA CHIMICA ACTA
卷 504, 期 1, 页码 73-79

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/S0003-2670(03)00673-1

关键词

propofol; biosensor; molecular imprinting; anaesthetic

向作者/读者索取更多资源

Herein we describe a new thin film imprinted polymer for propofol (active ingredient of a commonly used general intravenous anaesthetic) developed on a membrane support. Propofol binds to this polymer in 2 min and can be removed and tested in 1 min, giving a total assay time of 3 min. The non-specific binding to the polymer is below the detection limit (0.1 mug/ml) and the response rate is below the rate of metabolism of the anaesthetic. Tests performed with this polymer against propofol-spiked blood samples showed good linearity and specificity at clinically relevant concentrations of 1-10 mug/ml and the working range for the system is 0.1-50 mug/ml. The polymer is easily regenerated and can be re-used for subsequent testing (in blood up to 5000 cycles). These results suggest suitability for use in an on-line propofol detection system. (C) 2003 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据