4.6 Article

In vivo human retinal imaging by ultrahigh-speed spectral domain optical coherence tomography

期刊

OPTICS LETTERS
卷 29, 期 5, 页码 480-482

出版社

OPTICAL SOC AMER
DOI: 10.1364/OL.29.000480

关键词

-

类别

资金

  1. NEI NIH HHS [1R24 EY12877] Funding Source: Medline

向作者/读者索取更多资源

An ultrahigh-speed spectral domain optical coherence tomography (SD-OCT) system is presented that achieves acquisition rates of 29,300 depth profiles/s. The sensitivity of SD-OCT and time domain OCT (TD-OCT) are experimentally compared, demonstrating a 21.7-dB improvement of SD-OCT over TD-OCT. In. vivo images of the human retina are presented, demonstrating the ability to acquire high-quality structural images with an axial resolution of 6 mum at ultrahigh speed and with an ocular exposure level of less than 600 muW. (C) 2004 Optical Society of America.