4.7 Article

Standardization of immunoassays for measurement of myoglobin in serum. Phase 1: Evaluation of candidate secondary reference materials

期刊

CLINICA CHIMICA ACTA
卷 341, 期 1-2, 页码 65-72

出版社

ELSEVIER
DOI: 10.1016/j.cccn.2003.10.030

关键词

myoglobin; standardization; commutability; reference materials

向作者/读者索取更多资源

Background: Myoglobin is a low-molecular weight protein present in the cytosol of striated muscles. Its concentrations in serum can be measured by immunoassays and are used as an early indicator of myocardial necrosis. Since variability among commercial myoglobin assays exists, standardization of myoglobin assays is needed. Methods: An international collaborative study was organized with the involvement of seven companies using 12 different automated platforms for measuring myoglobin. Five candidate secondary, i.e., matrixed, reference materials were assayed in relation to linearity, imprecision, recovery rate and commutability to demonstrate a possible identity between the materials and the usual routine serum samples. Results: One Iyophilized candidate material (human heart myoglobin in human serum) was selected as the most suitable secondary reference material, based on the criteria examined. Used as a calibrator a posteriori, the bias between the various myoglobin assays for a frozen human serum pool was reduced from 32% to 13%. Conclusion: This study provides the basis for the selection of an internationally recognized secondary reference material. (C) 2004 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据