4.5 Article

Gated fiber-optic-coupled detector for in vivo real-time radiation dosimetry

期刊

APPLIED OPTICS
卷 43, 期 8, 页码 1663-1668

出版社

OPTICAL SOC AMER
DOI: 10.1364/AO.43.001663

关键词

-

类别

向作者/读者索取更多资源

Gated detection of the output of a fiber-optic-coupled radiation dosimeter effectively eliminated the direct contribution of Cerenkov radiation to the signal. The radiation source was an external beam radiotherapy machine that provided pulses of 6-MeV x rays. Gated detection was used to discriminate the signal collected during the radiation pulses, including Cerenkov interference, from the signal collected between the radiation pulses due only to phosphorescence from the Cu1+-doped glass detector. Gated detection of the long-lived phosphorescence of the Cu1+-doped glass provided real-time dose measurements that were linear with the absorbed dose and that were accurate for all field sizes studied. (C) 2004 Optical Society of America.