Diagnostic methods for time-resolved optical spectroscopy of shocked liquid deuterium

期刊

REVIEW OF SCIENTIFIC INSTRUMENTS
卷 75, 期 4, 页码 928-935

出版社

AMER INST PHYSICS
DOI: 10.1063/1.1651632

关键词

-

向作者/读者索取更多资源

Sandia National Laboratories Z facility generates shocks in liquid deuterium with pressures up to 100 GPa. Temperature measurements using spectroscopy of the shocked D-2 self-emission can help discriminate between different deuterium equation of state models. Time-resolved spectra are recorded using four diagnostic systems, each composed of a fiber optic probe that transmits light from the shocked D-2 to a remote streaked spectrometer. Calibration of the entire system in the streaked mode is performed using a xenon arc lamp. The absolute xenon arc lamp spectrum is determined by comparison with National Institute of Standards and Technology (NIST) standards. Data analysis is performed by measuring the wavelength-dependent efficiency for each system and applying this to determine the shocked D-2 self-emission spectrum. Temperature is deduced from either the wavelength dependence of the spectral radiance, ignoring the absolute intensity, or from both the wavelength dependence and the absolute intensity. The shocked D-2 temperature uncertainties obtained with the methods described here are well below the differences between D-2 equation of state (EOS) models and these measurements should therefore provide new insight into D-2 EOS physics. (C) 2004 American Institute of Physics.