4.6 Article

Accuracy of electrical aerosol sensors measuring lung deposited surface area concentrations

期刊

JOURNAL OF AEROSOL SCIENCE
卷 89, 期 -, 页码 96-109

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.jaerosci.2015.07.003

关键词

Lung deposited surface area concentration; Accuracy; Diffusion chargers; Monodisperse particles

资金

  1. French National Funding Agency for Research (ANR)
  2. German Federal Ministry of Education and Research (BMBF) [03X0127A]
  3. British Technology Strategy Board (TSB)
  4. Swiss TEMAS AG

向作者/读者索取更多资源

The accuracy of instruments measuring the alveolar lung deposited surface area (LDSA) concentration based on unipolar diffusion charging has been assessed. Monoclisperse particles with sizes between 10 nm and 700 nm were used The results indicate that the LDSA concentration can be measured with at least +/- 30% accuracy for particle sizes between 20 nm and 400 nm. The LDSA concentrations of particles < 20 nm are overestimated and of particles >400 nm increasingly underestimated by the instruments. The LDSA concentration of 685 nm particles was on average underreported by a factor of approximately 2. LDSA concentrations measured with agglomerated particles were consistently higher than those measured with spherical particles of the same electrical mobility diameter. The accuracy was, however, still mostly within +/- 30% for a size range from 20 nm to 400 nm. Due to the strong and increasing deviation for large particles, the use of an appropriate preseprator is recommended when measuring LDSA concentrations. (C) 2015 Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据