Real-time cardiac MRI at 3 tesla

期刊

MAGNETIC RESONANCE IN MEDICINE
卷 51, 期 4, 页码 655-660

出版社

JOHN WILEY & SONS INC
DOI: 10.1002/mrm.20053

关键词

real-time MRI; high field; cardiac imaging; spirals; spectral-spatial excitation

向作者/读者索取更多资源

Real-time cardiac and coronary MRI at 1.5T is relatively signal starved and the 3T platform is attractive for its immediate factor of two increase in magnetization. Cardiac imaging at 3T, however, is both subtly and significantly different from imaging at 1.5T because of increased susceptibility artifacts, differences in tissue relaxation, and RF homogeneity issues. New RF excitation and pulse sequence designs are presented which deal with the fat-suppression requirements and off-resonance issues at 3T. Real-time cardiac imaging at 3T is demonstrated with high blood SNR, blood-myocardium CNR, resolution, and image quality, using new spectral-spatial RF pulses and fast spiral gradient echo pulse sequences. The proposed sequence achieves 1.5 mm in-plane resolution over a 20 cm FOV, with a 5.52 mm measured slice thickness and 32 dB of lipid suppression. Complete images are acquired every 120 ms and are reconstructed and displayed at 24 frames/sec using a sliding window. Results from healthy volunteers show improved image quality, a 53% improvement in blood SNR efficiency, and a 232% improvement in blood-myocardium CNR efficiency compared to 1.5T. (C) 2004 Wiley-Liss, Inc.