REVIEW OF SCIENTIFIC INSTRUMENTS
卷 75, 期 4, 页码 837-841出版社
AMER INST PHYSICS
DOI: 10.1063/1.1688442
关键词
-
We describe a method of preparing gold scanning tunneling microscopy (STM) tips by direct current electrochemical etching in concentrated HCl and ethanol solution. Gold tips with tip apex radius lower than 30 nm can be reproducibly prepared by this method. The influence of the solution composition, etching voltage on the surface structure, and sharpness has been investigated. These tips can be efficiently used for STM imaging, tip-enhanced Raman spectroscopy, and light emission investigations on the same sample. (C) 2004 American Institute of Physics.