期刊
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION
卷 85, 期 4, 页码 598-603出版社
W B SAUNDERS CO-ELSEVIER INC
DOI: 10.1016/j.apmr.2003.07.013
关键词
hip; rehabilitation; reliability and validity
Objective: To compare the inter- and intrarater reliability of a portable dynamometer anchoring station (DAS) to a handheld dynamometer (HHD). Design: Repeated-measures design. Setting: Human performance and movement analysis laboratory. Participants: Fifteen healthy participants, ages 23 to 44 years. Interventions: Not applicable. Main Outcome Measures: Three consecutive measures of peak bilateral isometric strength were obtained for hip abduction, extension, and flexion by 2 investigators by using the DAS and the HHD after a 1-hour rest period. This testing scenario was repeated I week later. Intraclass correlation coefficients (ICCs) were used to determine reliability. Results: Interrater ICCs of average peak strength ranged from .84 to .92 (hip flexors), .69 to .88 (hip abductors), and .56 to .80 (hip extensors). Intrarater ICCs ranged from .59 to .89 for tester A and from .72 to .89 for tester B using the DAS, and from .67 to .81 for the HHD across muscle groups. Conclusions: The DAS showed good intrarater reliability for hip flexion and abduction, whereas the HHD demonstrated higher reliability for hip extension. The results support the use of dynamometers that are quick and reliable and that reduce tester bias during hip strength assessment.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐