4.5 Article

Proposed role of embedded dentin chips for the clinical failure of nickel-titanium rotary instruments

期刊

JOURNAL OF ENDODONTICS
卷 30, 期 5, 页码 339-341

出版社

ELSEVIER SCIENCE INC
DOI: 10.1097/00004770-200405000-00008

关键词

-

向作者/读者索取更多资源

Discarded ProFile and ProTaper nickel-titanium rotary instruments, with unknown history of clinical use, were obtained from graduate endodontic clinics at Ohio State University and University of Texas Health Science Center at Houston Dental Branch. These discarded instruments and as-received instruments of both types were examined with a scanning electron microscope to investigate effects of clinical use and causes of failure. For used ProTaper instruments, dentinal debris was wedged mostly in narrow, radial, land-type regions and less on convex flute surfaces. For used ProFile instruments, dentinal debris was wedged mostly in the metal rollover and on concave flute surfaces. Used instruments of both types exhibited widened machining grooves, and elongated and stretched rollover. Dentin chips were wedged in surface micro-cracks that appeared to propagate from original machining flaws and widen during in vivo root canal preparation. From our observational study, wedged dentinal deposits seem to play a pivotal role for clinical failure of these instruments.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据