4.8 Article

Characterization of a focal plane camera fitted to a Mattauch-Herzog geometry mass spectrograph. 2. Use with an inductively coupled plasma

期刊

ANALYTICAL CHEMISTRY
卷 76, 期 9, 页码 2531-2536

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac030337u

关键词

-

向作者/读者索取更多资源

A novel charge-sensitive detector array, termed the focal plane camera (FPC), has been coupled to a Mattauch-Herzog mass spectrograph (MHMS) with an inductively coupled plasma ionization source. The FPC employs an array of gold Faraday cups, each with its own charge-integrating circuit that allows the simultaneous detection of several m/z ratios. The ion-sampling interface of the MHMS has been redesigned to provide better heat transfer away from the sampler and skimmer cones and to reduce the negative effects of turbulent gas flows around the plasma. The instrument has produced limits of detection in the tens to hundreds of parts per quadrillion regime and isotope ratio accuracy and precision of 5% error and 0.007% RSD, respectively. Limits of detection with the FPC are comparable to those obtained with a single-channel secondary electron multiplier (SEM). However, the isotope ratio accuracy and precision are better with the FPC than when the SEM is employed. The dynamic range has been shown to be linear over 7 orders of magnitude.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据