A new smoothing procedure to reduce delivery segments for static MLC-based IMRT planning

期刊

MEDICAL PHYSICS
卷 31, 期 5, 页码 1158-1165

出版社

AMER ASSOC PHYSICISTS MEDICINE AMER INST PHYSICS
DOI: 10.1118/1.1713279

关键词

-

向作者/读者索取更多资源

In the application of pixel-based intensity-modulated radiation therapy (IMRT) using the step-and-shoot delivery method, one major difficulty is the prolonged delivery time. In this study, we present an integrated IMRT planning system that involves a simple smoothing method to reduce the complexity of the beam profiles. The system consists of three main steps: (a) an inverse planning process based on a least-square dose-based cost function; (b) smoothing of the intensity maps; (c) reoptimization of the segment weights. Step (a) obtains the best plan with the lowest cost value using a simulated annealing optimization algorithm with discrete intensity levels. Step (b) takes the intensity maps obtained from (a) and reduces the complexity of the maps by smoothing the adjacent beamlet intensities. During this process each beamlet is assigned a structure index based on anatomical information. A smoothing update is applied to average adjacent beamlets with the same index. To control the quality of the plan, a predefined clinical protocol is used as an acceptance criterion. The smoothing updates that violate the criterion are rejected. After the smoothing process, the segment weights are reoptimized in step (c) to further improve the plan quality. Three clinical cases were studied using this system: a medulloblastoma, a prostate cancer, and an oropharyngeal carcinoma. While the final plans demonstrate a degradation of the original plan quality, they still meet the plan acceptance criterion. On the other hand, the segment numbers or delivery times are reduced by 40%, 20%, and 20% for the three cases, respectively. (C) 2004 American Association of Physicists in Medicine.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据