4.2 Article

Online motion correction for diffusion-weighted segmented-EPI and FLASH imaging

出版社

SPRINGER
DOI: 10.1007/s10334-004-0036-x

关键词

diffusion magnetic; resonance imaging; online motion correction; navigator echoes; echo-planar imaging; TurboFLASH

向作者/读者索取更多资源

This paper explores the application of online motion correction using navigator echoes to the segmented-EPI and FLASH techniques. In segmented EPI this has the advantage over post-acquisition correction that the position in k-space of each segment is no longer subject to arbitrary shifts caused by rotation. In diffusion-weighted FLASH it has the advantage that the full magnetisation can be utilised in comparison to other methods of eliminating the sensitivity to bulk motion, in which the sensitivity is halved. Healthy subjects were investigated on a 3 T whole-body system in which the hardware has been modified so that navigator echoes can be recorded on a personal computer which generates the necessary magnetic field gradient correction pulses and shifts in the Larmor frequency within 800 mus. ECG triggering was used to avoid the period of non-rigid-body brain motion. Two orthogonal navigator echoes were employed. For segmented EPI it was found essential to minimise the T2* weighting of the navigator echoes to about 10 ms to obtain reliable results. High quality images were obtained for both methods examined. Online motion correction brings direct benefits to both the diffusion-weighted segmented-EPI and FLASH techniques.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据