4.3 Article Proceedings Paper

Comparative estimation of anterior chamber depth by ultrasonography, Orbscan II, and IOLMaster

期刊

JOURNAL OF CATARACT AND REFRACTIVE SURGERY
卷 30, 期 6, 页码 1268-1271

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.jcrs.2003.11.053

关键词

-

向作者/读者索取更多资源

Purpose: To assess the degree of agreement of anterior chamber depth (ACD) measurements by 2 optical devices (Orbscan II and IOLMaster) and compare them with contact ultrasound A-scan. Setting: Department of Ophthalmology at Hull Royal Infirmary, England. Methods: In this prospective study of 81 eyes of 41 consecutive patients, ACID estimation was done by 3 methods-scanning slit topography (Orbscan II), partial coherence interferometry (IOLMaster), and contact ultrasound A-scan in that order. The same observer performed all measurements. Repeated measures Analysis of Variance (ANOVA) was used to analyze difference between ACID measurements by the 3 methods. Difference in measurements between methods was assessed using the paired t test. Results: The mean ACD (+/-SD) by the 3 methods was 3.32 mm (0.60), 3.33 mm (0.61), and 2.87 mm (0.55), respectively. There was a statistically significant difference between measurements recorded by the 3 methods (P<.01). Mean contact A-scan measurements were 0.40 mm and 0.43 mm lower than by Orbscan II and IOLMaster respectively (P<.01). Conclusion: Applanation ultrasound gives consistently lower measurements for ACD compared to Orbscan If and IOLMaster. Although a high degree of agreement between Orbscan II and IOLMaster was noted, further studies are needed to assess the interchangeability of measurements in clinical practice.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据