4.6 Article

Endobronchial ultrasound to assess airway wall thickening:: validation in vitro and in vivo

期刊

EUROPEAN RESPIRATORY JOURNAL
卷 23, 期 6, 页码 813-817

出版社

EUROPEAN RESPIRATORY SOC JOURNALS LTD
DOI: 10.1183/09031936.04.00119904

关键词

airway wall measurements; computed tomography; endobronchial ultrasound

向作者/读者索取更多资源

Endobronchial ultrasound (EBUS) allows identification of airway wall structures and could potentially be utilised for in vivo studies of airway thickening in asthma. The present study investigated whether inflation of the fluid-filled balloon sheath over the transducer (necessary to provide sonic coupling with the airway wall) influenced in vitro measurements. In vivo comparability of EBUS with high resolution computed tomography scanning (HRCT), an established method for measuring wall thickness, was determined in control subjects. The airway diameter and wall thickness were studied using EBUS in 24 cartilaginous airways obtained from four sheep, before and after balloon sheath inflation during immersion in saline. To assess EBUS versus HRCT comparability of airway measures in vivo, 12 control subjects underwent imaging of the posterior basal bronchus of the right lower lobe by both techniques. Intra- and interobserver agreement were also assessed. Results with and without the balloon sheath gave comparable measures of airway internal diameter and wall thickness in vitro. Statistical analysis showed agreement between EBUS and HRCT, and intra- and interobserver variability in vivo. The current study concludes that endobronchial ultrasound, which does not present a radiation risk, could be utilised in the in vivo study of cartilaginous airway wall remodelling in respiratory diseases, such as asthma.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据