4.8 Article Proceedings Paper

Amperometric detection of morphine at a Prussian blue-modified indium tin oxide electrode

期刊

BIOSENSORS & BIOELECTRONICS
卷 20, 期 1, 页码 3-8

出版社

ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/j.bios.2003.11.027

关键词

amperometric detection; electrochemical sensor; artificial peroxidase; morphine; Prussian blue

向作者/读者索取更多资源

In this work, the electrocatalytic oxidation of morphine (MO) at an optically transparent indiurn tin oxide (ITO) electrode modified by an electrodeposited Prussian blue (PB) thin film is first demonstrated, and the amperometric detection of MO was then investigated. Experimental results showed that the thin film on the ITO surface, confined to the PB/Berlin green (BG) redox pair, can serve as an excellent mediator which facilitates electron transfer and considerably lowers the overpotential required, as compared to a bare ITO electrode. Thus, PB can be regarded as a promising artificial peroxidase for MO. The rate of such an electrocatalytic reaction is pH dependent with the highest value at pH 5. By potential-step excitation from 0.55 to 0.70 V, a linear calibration curve, displaying the relationship between steady-state currents and MO concentrations (ranging from 0.09 to 1.0 mM), was obtained. The detection sensitivity is about 16.8 muA/cm(2) mM. Most importantly, the method described herein can readily discriminate MO analogs lacking the phenolic -OH group, such as codeine, and can thus benefit the specific recognition of MO. (C) 2004 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据