Evaluation of specific absorption rate as a dosimeter of MRI-related implant heating

期刊

JOURNAL OF MAGNETIC RESONANCE IMAGING
卷 20, 期 2, 页码 315-320

出版社

JOHN WILEY & SONS INC
DOI: 10.1002/jmri.20103

关键词

magnetic resonance imaging; bioeffects; magnetic resonance imaging; safety; neurostimulation systems; implants; deep brain stimulation; specific absorption rate

资金

  1. NINDS NIH HHS [NS 44575-01] Funding Source: Medline

向作者/读者索取更多资源

Purpose: To compare the magnetic resonance imaging (MRI)-related heating per unit of whole body averaged specific absorption rate (SAR) of a conductive implant exposed to two different 1.5-Tesla/64 MHz MR systems. Materials and Methods: Temperature changes at the electrode contacts of a deep brain stimulation lead were measured using fluoroptic thermometry. The leads were placed in a typical surgical implant configuration within a gel-filled phantom of the human head and torso. MRI was performed using two different transmit/receive body coils on two different generation 1.5-Tesla MR systems from the same manufacturer. Temperature changes were normalized to whole body averaged SAR values and compared between the two scanners. Results: Depending on the landmark location, the normalized temperature change for the implant was significantly higher on one MR system compared to the other (P<0.001). Conclusion: The findings revealed marked differences across two MR systems in the level of radiofrequency (RF)induced temperature changes per unit of whole body SAR for a conductive implant. Thus, these data suggest that using SAR to guide MR safety recommendations for neurostimulation systems or other similar implants across different MR systems is unreliable and, therefore, potentially dangerous. Better, more universal, measures are required in order to ensure patient safety.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据