3.9 Article

Two simple and generic antibody-independent kinase assays: Comparison of a bioluminescent and a microfluidic assay format

期刊

JOURNAL OF BIOMOLECULAR SCREENING
卷 9, 期 5, 页码 409-416

出版社

SAGE PUBLICATIONS INC
DOI: 10.1177/1087057104264175

关键词

HTS; serine/threonine kinase assay; bioluminescence; luciferase; microfluidics; caliper

向作者/读者索取更多资源

In this study, the authors have compared the performance of 2 high-throughput screening assays for a serin/threonine kinase: a microplate-based, bioluminescent assay that uses the luciferin/luciferase system to monitor ATP consumption, and a microfluidic assay that measures the change in mobility in an electric field of a fluorescently labeled peptide upon phosphorylation. Both assays are homogeneous, nonradioactive, antibody independent and could be miniaturized to a reaction volume of 4 mul. The robustness of both formats was demonstrated by Z' values > 0.8. Screening of a small library (2133 compounds) showed that the results obtained with both technologies correlate very well. Although the threshold for hits was set to a comparably low value-22.2% and 13.7% inhibition for the ATP consumption and microfluidic assay, respectively, corresponding to mean plus 3 standard deviations-the overlap of active compounds identified with the 2 assay formats was greater than 94%. Thus, both assays allow the identification of even low potency inhibitors with a high level of confidence.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.9
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据