The accuracy of free-hand cup positioning -: a CT based measurement of cup placement in 105 total hip arthroplasties

期刊

INTERNATIONAL ORTHOPAEDICS
卷 28, 期 4, 页码 198-201

出版社

SPRINGER
DOI: 10.1007/s00264-004-0542-5

关键词

-

向作者/读者索取更多资源

We studied 105 patients who received a total hip arthroplasty between June 1985 and August 2001 using freehand positioning of the acetabular cup. Using pelvic CT scan and the hip-plan module of SurgiGATE-System (Medivision, Oberdorf, Switzerland), we measured the angles of inclination and anteversion of the cup. Mean inclination angle was 45.8degrees+/-10.1degrees (range: 23.0-71.5degrees) and mean anteversion angle was 27.3degrees+/-15.0degrees (range: -23.5degrees to 59.0degrees). We compared the results to the safe position as defined by Lewinnek et al. and found that only 27/105 cups were implanted within the limits of the safe position. We conclude that a safe position as defined by Lewinnek et al. [13] was only achieved in a minority of the cups that were implanted freehand.