4.6 Article

Accurate Wavelength Calibration Method for Flat-Field Grating Spectrometers

期刊

APPLIED SPECTROSCOPY
卷 65, 期 9, 页码 1083-1086

出版社

SOC APPLIED SPECTROSCOPY
DOI: 10.1366/11-06280

关键词

Wavelength calibration; Parameter fitting; Spectrometer alignment; Flat-field spectrometers; Spherical varied line space gratings

资金

  1. National High Technology Research and Development Program (863 Program) of China [2010AA10Z201]
  2. Chinese Academy of China [YZ200922]

向作者/读者索取更多资源

A portable spectrometer prototype is built to study wavelength calibration for flat-field grating spectrometers. An accurate calibration method called parameter fitting is presented. Both optical and structural parameters of the spectrometer are included in the wavelength calibration model, which accurately describes the relationship between wavelength and pixel position. Along with higher calibration accuracy, the proposed calibration method can provide information about errors in the installation of the optical components, which will be helpful for spectrometer alignment.