Continuum Source Tungsten Coil Atomic Fluorescence Spectrometry

期刊

APPLIED SPECTROSCOPY
卷 65, 期 4, 页码 382-385

出版社

SAGE PUBLICATIONS INC
DOI: 10.1366/10-06158

关键词

Electrothermal vaporizer; Tungsten coil; Atomic fluorescence spectrometry; AFS; Continuum source; Atomic emission spectrometry; AES; Portable instrumentation

资金

  1. National Science Foundation
  2. Department of Homeland Security [NSF CBET 0736214, DHS 2008-DN-077-ARI020-03]

向作者/读者索取更多资源

A simple continuum source tungsten coil atomic fluorescence spectrometer is constructed and evaluated. The heart of the system is the atomizer: a low-cost tungsten filament extracted from a 150 W light bulb. The filament is resistively heated with a small, solid-state, constant-current power supply. The atomizer is housed in a glass chamber and purged with a 1 L/min flow of a conventional welding gas mixture: 10% H-2/Ar. A 25 mu L sample aliquot is pipetted onto the tungsten coil, the liquid is dried at low current, and then the atomic vapor is produced by applying a current in the range 3.5-5.5 A. The atomization current does not produce temperatures high enough to excite atomic emission. Radiation from a 300 W xenon lamp is focused through the atomic vapor, exciting atomic fluorescence. Fluorescence signals are collected using a hand-held charge-coupled device (CCD) spectrometer. Simultaneous determination of ten elements (Ag, Bi, Cr, Cu, Ga, In, Mg, Mn, and TI) results in detection limits in the range 0.3 to 10 ng. The application of higher atomization currents (10 A) leads to straightforward detection of atomic emission signals with no modifications to the instrument.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据