AMERICAN JOURNAL OF INFECTION CONTROL
卷 32, 期 5, 页码 291-295出版社
MOSBY-ELSEVIER
DOI: 10.1016/j.ajic.2003.12.004
关键词
-
Objective: The aim of this study was to assess the efficacy of a disinfectable, needle-free connector in the prophylaxis of catheter-related bloodstream infection. Methods: A randomized controlled trial was performed in a polyvalent intensive care unit. Patients who needed multilumen central venous catheters were randomly assigned to a study or a control group. All catheters were inserted and manipulated according to the Centers for Disease Control and Prevention (CDC) recommendations. Study group patients were equipped with catheters with disinfectable, needle-free connectors whereas control group patients were equipped with catheters with 3-way stopcocks. Two peripheral blood cultures and a semiquantitative culture of the catheter tip were performed on removal of the catheter. Results: The study included 243 patients, with a total of 278 central venous catheters. The catheters' mean insertion duration was 9.9 days. Both groups were comparable regarding patient and catheter characteristics. Incidence rate of catheter-related bloodstream infection was 0.7 per 1000 days of catheter use in the study group, compared with 5.0 per 1000 days of catheter use in the control group (P = .03). Conclusions: To add a disinfectable, needle-free connector to the CDC recommendations reduces the incidence of catheter-related bloodstream infection in critically ill patients with central venous catheters.