期刊
APPLIED SPECTROSCOPY
卷 62, 期 11, 页码 1180-1188出版社
SAGE PUBLICATIONS INC
DOI: 10.1366/000370208786401581
关键词
Line-scanning; Raman mapping; Raman microspectroscopy; Metamorphism; Carbonates
资金
- INSU DyETI program
- ANR Jeunes Chercheurs program
By allowing nondestructive chemical and structural imaging of heterogeneous samples with a micrometer spatial resolution, Raman mapping offers unique capabilities for assessing the spatial distribution of both mineral and organic phases within geological samples. Recently developed line-scanning Raman mapping techniques have made it possible to acquire Raman maps over large, millimeter-sized, zones of interest owing to a drastic decrease of the data acquisition time without losing spatial or spectral resolution. The synchronization of charge-coupled device (CCD) measurements with x,y motorized stage displacement has allowed dynamic line-scanning Raman mapping to be even more efficient: total acquisition time may be reduced by a factor higher than 100 compared to point-by-point mapping. Using two chemically and texturally complex geological samples, a fossil megaspore in a metamorphic rock and aragonite-garnet intergrowths in an Eclogitic marble, we compare here two recent versions of line-scanning Raman mapping systems and discuss their respective advantages and disadvantages in terms of acquisition time, image quality, spatial and imaging resolutions, and signal-to-noise ratio. We show that line-scanning Raman mapping techniques are particularly suitable for the characterization of such samples, which are representative of the general complexity of geological samples.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐