An integrated microfluidic device has been fabricated for analyzing the chemical contents of a single cell (chemical cytometry). The device is designed to accomplish four different functions: (i) cell handling, (it) metering and delivering of chemical reagents, (iii) cell lysis and chemical derivatization, and (iv) separating derivatized compounds and detecting them by laser-induced fluorescence. These functions are accomplished with only two valves, formed by multilayer soft lithography. A new kind of three-state valve and a picopipette are described; these elements are crucial for minimizing the reaction volume and ensuring optimal shape of the channel for electrophoresis injection. By using these valves, a reaction volume of approximate to70 pl is achieved for the lysis and derivitization of the contents of a single Jurkat T cell (approximate to10 mum diameter). As a demonstration of the use of this integrated microfluidic device, electropherograms of amino acids from individual Jurkat T cells are recorded and compared with those collected from a multiple-cell homogenate.