4.3 Article

Failure of less invasive stabilization system plates in the distal femur - A report of four cases

期刊

JOURNAL OF ORTHOPAEDIC TRAUMA
卷 18, 期 8, 页码 565-570

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/00005131-200409000-00015

关键词

less invasive stabilization system; LISS; failure; distal femur fractures

向作者/读者索取更多资源

Four cases of failure of distal femoral Less Invasive Stabilization System plates are reported. Two modes of failure were observed: in 2 cases, the Less Invasive Stabilization System plates broke, and in 2 cases, the plates gradually displaced from the femur as the shaft screws lost purchase in the diaphyseal femoral cortex. Delayed union/nonunion, early weight bearing, and improper plate placement were identified as the etiologies for plate failure.