JOURNAL OF MAGNETIC RESONANCE
卷 170, 期 1, 页码 79-87出版社
ACADEMIC PRESS INC ELSEVIER SCIENCE
DOI: 10.1016/j.jmr.2004.06.001
关键词
MRI; low-field scanner; food; phase shift; thermometry
MRI thermometry methods are usually based on the temperature dependence of the proton resonance frequency. Unfortunately, these methods are very sensitive to the phase drift induced by the instability of the scanner which prevents any temperature mapping over long periods of time. A general method based on 3D spatial modelling of the phase drift as a function of time is presented. The MRI temperature measurements were validated on gel samples with uniform and constant temperature and with a linear temperature gradient. In the case of uniform temperature conditions, correction of the phase drift proved to be essential when long periods of acquisition were required, as bias could reach values of up to 200 degreesC in its absence. The temperature uncertainty measured by MRI was 1.2 degreesC in average over 290 min. This accuracy is coherent with the requirements for food applications especially when thermocouples are useless. (C) 2004 Elsevier Inc. All rights reserved.