APPLIED RADIATION AND ISOTOPES
卷 67, 期 5, 页码 706-710出版社
PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.apradiso.2009.01.015
关键词
Gamma spectrometry; Detection efficiency; Ge detector; Monte Carlo simulation; Geant4; Optimised detector model
An optimisation of the geometrical model of a p-type detector used for material screening was carried out to improve the accuracy of Monte Carlo simulations in reproducing spectrometric measurements. Gamma-ray sources were measured to determine the dimensions of the detector dead layer and borehole. An agreement between simulations and measurement within 3% was achieved at energies above 100 keV. In contrast, discrepancies on the order of 23% were encountered using the nominal parameters from the detector manufacturer. (c) 2009 Elsevier Ltd. All rights reserved.