Pencil-beam vs fan-beam dual-energy X-ray absorptiometry comparisons across four systems

期刊

JOURNAL OF CLINICAL DENSITOMETRY
卷 7, 期 3, 页码 281-289

出版社

ELSEVIER SCIENCE INC
DOI: 10.1385/JCD:7:3:281

关键词

dual-energy X-ray absorptiometry; pencil-beam DXA; fan-beam DXA; bone density; body fat; body composition

资金

  1. PHS HHS [P01-42618] Funding Source: Medline

向作者/读者索取更多资源

We compared bone mineral density (BMD) and body fat percentage (%fat) between two pencil-beam (Lunar DPX and DPX-L) and two fan-beam (Lunar Prodigy, Hologic DelphiA) dual-energy X-ray absorptiometry (DXA) systems. We examined these values in the total-body, spine, femur, and forearm scans in 78 healthy adults across these four DXA systems. BMD and %fat values were highly correlated among the four instruments. DPX-L gave the lowest mean %fat and Prodigy gave the highest mean %fat for both sexes. The means were system dependent for %fat estimates across the four DXA machines. There was a significant difference detected in BMD estimates across manufacturers, with the DelphiA providing systematically lower values than the Lunar systems in the whole body, spine, and femur sites but higher values than the Lunar systems in the forearm. The present study results show that both %fat and bone mineral estimates between pencil-beam and fan-beam systems are highly correlated, but vary by system. Significant differences exist between the instruments, especially between different manufacturers, and most of the comparisons are sex dependent. We conclude that longitudinal studies should always be evaluated on the same system when possible, and translation models should be used to assess cross-instrument differences.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.0
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据