卷 64, 期 1, 页码 258-263出版社
ELSEVIER SCIENCE BV
DOI: 10.1016/j.talanta.2004.02.017
关键词
organolead compounds; vinegar; lCP-MS
Lead concentrations of 59 different types of vinegars (15-307 mug l(-1) in balsamic vinegars and 36-50 mug l(-1) in wine vinegars) were determined using both inductively coupled plasma mass spectrometry (ICP-MS) and graphite furnace atomic absorption spectrometry (GFAAS). Although the precision of direct analyses, following simple aqueous dilutions, with either instrumental method was poor; that precision, following nitric acid and/or hydrogen peroxide digestions, markedly improved with either instrument and the values obtained with the two instruments were in good agreement. The efficacy of different digestions, including (1) nitric acid using a heating block, with or without addition of hydrogen peroxide and (2) mixtures of nitric acid and hydrogen peroxide using ultraviolet (UV) photolysis, were then assessed. The latter procedure was found to be much faster and more efficient, but it was limited by the relatively high levels of contaminant lead in hydrogen peroxide. Consequently, it is recommended that lead concentrations in vinegar be measured following a nitric acid digestion and UV photolysis to oxidize all organic matter before ICP-MS or GFAAS analysis; and it is further recommend that the thermal settings for the latter analyses be adjusted to account for the apparent presence of relatively volatile organolead compounds in vinegar digests. (C) 2004 Elsevier B.V. All rights reserved.