期刊
ELECTROANALYSIS
卷 16, 期 20, 页码 1697-1703出版社
WILEY-V C H VERLAG GMBH
DOI: 10.1002/elan.200303016
关键词
multi-walled carbon nanotubes; glassy carbon; glucose oxidase; poly(phenylenediamine); amperometric biosensor
An amperometric glucose biosensor based on multi-walled carbon nanotube (MWCNT) modified glassy carbon electrode has been developed. MWCNT-modified glassy carbon electrode was obtained by casting the electrode surface with multi-walled carbon nanotube materials. Glucose oxidase was co-immobilized on the MWCNT-modified glassy carbon surface by electrochemical deposition of poly (o-phenylenediamine) film. Enhanced catalytic electro-reduction behavior of oxygen at MWCNT-modified electrode surface was observed at a potential of -0.40 V (vs. Ag \ AgCl) in neutral medium. The steady-state amperometric response to glucose was determined at a selected potential of -0.30 V by means of the reduction of dissolved oxygen consumed by the enzymatic reaction. Common interferents such as ascorbic acid, 4-acetamidophenol, and uric acid did not interfere in the glucose determination. The linear range for glucose determination extended to 2.0 mM and the detection limit was estimated to be about 0.03 mM.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐