4.5 Article

Coupled-column liquid chromatography method with photochemically induced derivatization for the direct determination of benzoylureas in vegetables

期刊

JOURNAL OF SEPARATION SCIENCE
卷 27, 期 14, 页码 1173-1180

出版社

WILEY-V C H VERLAG GMBH
DOI: 10.1002/jssc.200301661

关键词

benzoylurea insecticides; photochemically induced fluorescence; LC-LC separation; vegetables

向作者/读者索取更多资源

The coupled-column (LC-LC) system, consisting of a first column packed with internal surface reversed phase (ISRP) (50 x 4.6 mm ID) and a Chrompack C-18 (100 x 4.6 mm ID) as second column, allowed the simultaneous determination of five benzoylurea insecticides in dichloromethane (CH2Cl2) extracts of vegetable samples without any clean-up step. This system was combined with a photochemically induced fluorescence (PIF) post-column derivatization in order to provide strongly fluorescent photoproducts from the non-fluorescent benzoylureas. Limits of detection ranged from 0.21 to 0.98 mug L-1 of pesticide (equivalent to 0.14-0.65 mug kg(-1) in vegetable samples) and limits of determination ranged from 4.0 to 10.0 mug L-1 (equivalent to 2.7-6.7 mug kg(-1)). Linearity of the method was established between 2 and 1800 mug L-1, depending upon the compound. Validation of the total method was performed by randomly analyzing recoveries of four vegetable samples (aubergine, cucumber, green bean, and tomato) spiked at two levels of concentration (10.0 and 33.3 mug kg(-1)). The combination of the LC-LC system with PIF detection provides a sensitive, selective, and rapid method for the determination of pesticides in vegetable samples at levels lower than the maximum residue levels (MRLs) established for these compounds by Spanish legislation.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据