This study examined used, discarded rotary nickel-titanium instruments obtained from 14 endodontists in four countries, and identified factors that may influence defects produced during clinical use. A total of 7,159 instruments were examined for the presence of defects. Unwinding occurred in 12% of instruments and fractures in 5% (1.5% torsional, 3.5% flexural). The defect rates varied Significantly among endodontists. Instrument design factors also influenced defect rate, but to a lesser extent. The mean number of uses of instruments with and without defects was 3.3 +/- 1.8 (range: 1-10), and 4.5 +/- 2.0 (range: 1-16), respectively. The most important influence on defect rates was the operator, which may be related to clinical skill or a conscious decision to use instruments a specified number of times or until defects were evident.