A flow injection analysis (FIA) system has been developed for the flame atomic absorption spectroscopic (FAAS) determination of zinc in iron matrix. The spectral line interference of iron at 213.859 nm was eliminated by on-line separation using a micro-column of strong anion-exchange resin (Dowex I-M). The zinc chloro complexes were retained from 2 M HCl solution while most of the iron chloro complexes were passed to waste. For a 2% iron solution, matrix removal efficiency was 98.2% which means that positive spectral line interference of iron at the Zn line was reduced from 0.42 to 0.008 mug ml(-1) Zn. The optimized flow injection system can handle up to 48 samples with good precision (less than 3.5% relative standard deviation (R.S.D.)) in the working range of 0.075-2.2 mug ml(-1) Zn. Comparative analysis of a certified reference material and synthetic sample solutions containing traces of Zn in 2% Fe by the proposed method and by graphite furnace atomic absorption spectroscopy (GFAAS) showed no evidence of analytical bias at the 95% confidence level. (C) 2004 Elsevier B.V. All rights reserved.