4.6 Article

A carbon nanotube field emission multipixel x-ray array source for microradiotherapy application

期刊

APPLIED PHYSICS LETTERS
卷 98, 期 21, 页码 -

出版社

AMER INST PHYSICS
DOI: 10.1063/1.3595268

关键词

-

资金

  1. NCI [5R21-CA128510-01-03]
  2. NCI CCNE [U54-CA119343-01]

向作者/读者索取更多资源

The authors report a carbon nanotube (CNT) field emission multipixel x-ray array source for microradiotherapy for cancer research. The developed multipixel x-ray array source has 50 individually controllable pixels and it has several distinct advantages over other irradiation source including high-temporal resolution (millisecond level), the ability to electronically shape the form, and intensity distribution of the radiation fields. The x-ray array was generated by a CNT cathode array (5 x 10) chip with electron field emission. A dose rate on the order of >1.2 Gy/min per x-ray pixel beam is achieved at the center of the irradiated volume. The measured dose rate is in good agreement with the Monte Carlo simulation result. (C) 2011 American Institute of Physics. [doi: 10.1063/1.3595268]

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据