4.0 Article

Test-retest reliability of the multifocal electroretinogram and Humphrey visual fields in patients with retinitis pigmentosa

期刊

DOCUMENTA OPHTHALMOLOGICA
卷 109, 期 3, 页码 255-272

出版社

SPRINGER
DOI: 10.1007/s10633-005-0567-0

关键词

multifocal electroretinogram; reliability; retinitis pigmentosa; visual fields

向作者/读者索取更多资源

We examined the reliability of Humphrey visual field thresholds and multifocal electroretinogram (mfERG) amplitudes and timing in a group of patients with Retinitis Pigmentosa (RP). Eight patients with RP and seven control subjects were tested five times: at baseline (visit #0), at three weekly follow-up visits (visits #1 - #3), and at three months (visit #4). For the Humphrey thresholds, differences between dB values on repeat visits were obtained. Differences between log values on repeat visits were calculated for mfERG amplitude and implicit time. We used the standard deviations of these difference scores as a measure of reliability and the means of the difference scores as a measure of progression. We found that the majority of the patients' repeat data were more variable than that of the control subjects for both the Humphrey and mfERG. We found no single factor that predicted the magnitude, or the variance, of the SD of differences scores for the patients. We recommend that each patient's reliability be assessed individually. Ultimately, the choice of an outcome measure must be guided by its reliability, as well as its ability to assess the visual function of interest.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.0
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据