4.8 Article

Preparative linear ion trap mass spectrometer for separation and collection of purified proteins and peptides in arrays using ion soft landing

期刊

ANALYTICAL CHEMISTRY
卷 76, 期 21, 页码 6293-6305

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac048981b

关键词

-

向作者/读者索取更多资源

A preparative mass spectrometer for microarray fabrication is reported. The instrument includes an atmospheric pressure ionization source, a linear ion trap mass analyzer, an ion collection surface positioning system, and a surface loading chamber with independent vacuum pumping. It was designed for the production of protein arrays using the ion soft-landing technique to collect ions on a surface after separation by mass/charge ratio. Small microarrays have been prepared by isolating and soft landing individual protein or peptide ions after electrospray ionization of mixtures. The composition and purity of the separated materials has been confirmed using independent external mass spectrometric analysis of rinse solutions of the collected spots, either by the new method of electrosonic spray ionization MS or by nanospray ionization MS. The ability to retain bioactivity in the mass-selected and collected biomolecules has been demonstrated in particular cases. The reported instrument has also been characterized as an analytical mass spectrometer.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据