Geometric distortion in clinical MRI systems - Part II: correction using a 3D phantom

期刊

MAGNETIC RESONANCE IMAGING
卷 22, 期 9, 页码 1223-1232

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.mri.2004.08.014

关键词

geometric distortion; full restoration; piecewise interpolation; MRI; 3D phantom

向作者/读者索取更多资源

In this paper, we present the correction of the geometric distortion measured in the clinical magnetic resonance imaging (MRI) systems reported in the preceding paper (Part 1) using a 3D method based on the phantom-mapped geometric distortion data. This method allows the correction to be made on phantom images acquired without or with the vendor correction applied. With the vendor's 2D correction applied, the method corrects for both the residual geometric distortion still present in the plane in which the correction method was applied (the axial plane) and the uncorrected geometric distortion along the axis non-nal to the plane. The evaluation of the effectiveness of the correction using this new method was carried out through analyzing the residual geometric distortion in the corrected phantom images. The results show that the new method can restore the distorted images in 3D nearly to perfection. For all the MRI systems investigated, the mean absolute deviations in the positions of the control points (along x-, y- and z-axes) measured on the corrected phantom images were all less than 0.2 mm. The maximum absolute deviations were all below similar to0.8 mm. As expected, the correction of the phantom images acquired with the vendor's correction applied in the axial plane performed equally well. Both the geometric distortion still present in the axial plane after applying the vendor's correction and the uncorrected distortion along the z-axis have all been restored. (C) 2004 Elsevier Inc. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据