4.6 Article

Accurate height and volume measurements on soft samples with the atomic force microscope

期刊

LANGMUIR
卷 20, 期 23, 页码 10038-10045

出版社

AMER CHEMICAL SOC
DOI: 10.1021/la048650u

关键词

-

向作者/读者索取更多资源

The suitability of three common atomic force microscope (AFM) imaging modes for quantitative height and volume measurements on soft samples was investigated. The height and volume of rehydrated human metaphase chromosomes in liquid were measured using the contact mode, the tapping mode, and the force mapping mode. In both the contact and tapping modes, the measured height and volume strongly depended on the imaging setpoint that sets the imaging force. Measurement deviations up to 50% were observed. The force mapping mode, on the other hand, yielded reproducible height and volume measurements independent of the imaging force. It is therefore suggested that the force mapping mode should be used whenever the height or volume of soft samples need to be accurately determined.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据