期刊
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS
卷 36, 期 3, 页码 457-464出版社
ELSEVIER
DOI: 10.1016/j.jpba.2004.07.029
关键词
restricted access material; LC-MS/MS; direct injection; online extraction
We describe an automated approach to analyzing whole plasma samples using online extraction without the need for an analytical column. A single restricted access material (RAM) column provided online extraction and pre-concentration of analytes while effectively removing proteins. salts and other biological materials found in the plasma sample matrix. The reduction in the plasma matrix enabled direct elution of the analytes from the extraction column to the mass spectrometer for selective detection. The precision of the method was evaluated using a proprietary therapeutic agent (Compound A) and was less than 5% over the range of 1-500 ng/ml in spiked whole plasma, with an LOQ of 1 ng/ml. A side-by-side comparison of RAM results from a pharmacokinetic study in rats was made with a traditional protein precipitation LC-MS method and a correlation of 0.993 was obtained between both methods. The injection-to-injection cycle time for the RAM method was 8 min. Further automation was demonstrated by addition and mixing of the internal standard to all samples via an injection program of the autosampler. (C) 2004 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐