Matrix effect in LC-ESI-MS and LC-APCI-MS with off-line and on-line extraction procedures

期刊

JOURNAL OF CHROMATOGRAPHY A
卷 1058, 期 1-2, 页码 61-66

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.chroma.2004.08.118

关键词

matrix effects; methadone; off-line sample preparation; on-line solid-phase extraction

向作者/读者索取更多资源

Matrix effect on mass spectrometry response was investigated with commercially available electrospray ionisation (ESI) and atmospheric pressure chemical ionisation (APCI) sources coupled with a single quadruple mass spectrometer. A post-column infusion system was used to observe the MS signal alterations of methadone, selected as the model compound, in plasma. For this purpose, two sample preparation procedures were tested: (1) conventional off-line sample preparations with liquid-liquid extraction (LLE), solid-phase extraction (SPE) and protein precipitation (PP) with perchloric acid (PA) and acetonitrile (ACN) and (2) on-line SPE with two different extraction columns packed with a large particle support (LPS) and with restricted access material (RAM), respectively. Whatever the sample preparation procedures, APCI source appeared to be less liable to matrix effect than ESI source. Among the different off-line sample preparations, LLE was the most efficient extraction procedure. With other techniques, MS signal was affected not only by endogenous material but also by procedure introduced compounds. Moreover, on-line SPE-LC-MS configuration exhibited matrix effects, which depend on the API source and the extraction support. (C) 2004 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据