4.6 Article

Evaluation of nano-liquid chromatography-tandem mass spectrometry in a column switching setup for the absolute quantification of peptides in the picomolar range

期刊

JOURNAL OF CHROMATOGRAPHY A
卷 1058, 期 1-2, 页码 113-119

出版社

ELSEVIER
DOI: 10.1016/j.chroma.2004.07.098

关键词

neuropeptides; leu-enkephalin; quantification

向作者/读者索取更多资源

A standard nanospray-liquid chromatography-tandem mass spectrometry system in a column switching setup for the absolute quantification of leucine-enkephalin was evaluated. Analytes were loaded on a C18 trapping column and back-flushed in the 75 mum analytical column. Quantification was performed with a triple quadrupole instrument. Validation results show that it is feasible, with a conventional nano-LC system in the column switching setup, to quantify peptides as low as 500 amol on column (50 pmol/L). Weighted linear regression analysis proves a good linearity in a dynamic range of almost three orders of magnitude. Nevertheless, robustness remains a key issue in nano-LC-MS/MS. (C) 2004 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据