4.6 Article

Development and validation of a direct, non-destructive quantitative method for medroxyprogesterone acetate in a pharmaceutical suspension using FT-Raman spectroscopy

期刊

EUROPEAN JOURNAL OF PHARMACEUTICAL SCIENCES
卷 23, 期 4-5, 页码 355-362

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.ejps.2004.08.009

关键词

FT-Raman spectroscopy; HPLC; pharmaceutical suspension; validation

向作者/读者索取更多资源

A simple linear regression method was developed and statistically validated for the direct and non-destructive quantitative analysis-without sample preparation-of the active pharmaceutical ingredient (API) medroxyprogesterone acetate (MPA) in an aqueous pharmaceutical suspension (150 mg in 1.0 ml) using Fr-Raman spectroscopy. The linear regression was modelled by plotting the highest peak intensity of the vector normalized spectral band between 1630 and 1590 cm(-1) against different MPA standard suspension concentrations. At this band, no spectral interferences from additives in the suspension are observed. The validated model was used for the quantification of a commercial suspension (150 mg in 1.0 ml) of the commercialized preparations. The same standards and samples were used, respectively, for the development and validation of a simple linear regression model and for the quantitative determination by means of HPLC-with sample preparation-as described for the related substances of MPA in the Ph. Eur. IV. The quantification results obtained by the FT-Raman method corresponded with the claimed label concentration (150.01 +/- 0.96 mg/ml (n = 6)). Applying the HPLC method, however, a systematic error was observed (157.77 +/- 0.94 mg/ml (n = 6)). The direct FT-Raman method hence appears the most reliable for the quantification of the MPA component in suspension, compared to the HPLC method that requires sample preparation. The latter method provides a systematic error because the exact volume or density of a suspension sample is unknown. A precise isolation of fixed volumes from a suspension is rather unfeasible because of the continuous sagging of the suspended particles and their sticking to the used materials in the isolation process. (C) 2004 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据