JOURNAL OF CHROMATOGRAPHY A
卷 1062, 期 1, 页码 57-64出版社
ELSEVIER SCIENCE BV
DOI: 10.1016/j.chroma.2004.11.015
关键词
column selectivity; reversed-phase liquid chromatography; cyanopropyl columns
Eleven cyanopropyl (cyano) columns were characterized by means of a relationship developed originally for alkyl-silica columns. Compared to type-B alkyl-silica columns (i.e., made from pure silica), cyano columns are much less hydrophobic (smaller H), less sterically restricted (smaller S*), and have lower hydrogen-bond acidity (smaller A). Because sample retention is generally much weaker on cyano versus other columns (e.g., C-8, C-18), a change to a cyano column usually requires a significantly weaker mobile phase in order to maintain comparable values of k for both columns. For this reason, practical comparisons of selectivity between cyano and other columns (i.e., involving different mobile phases for each column) must take into account possible changes in separation due to the change in mobile phase, as well as change in the column. (C) 2004 Elsevier B.V. All rights reserved.