This report describes an optical sensing hybrid system obtained by bonding a microfluidic system to an integrated optical (10) four-channel Young interferometer (YI) chip. The microfluidic system implemented into a glass plate consists of four microchannels with cross-sectional dimensions of 200 mum x 15 mum. The microfluidic system is structured in such a way that after bonding to the IO chip, each microchannel addresses one sensing window in the four-channel YI sensor. Experimental tests show that the implementation of the microfluidics reduces the response time of the sensor from 100 s, as achieved with a bulky cuvette, to 4 s. Monitoring the anti-human serum albumine/human serum albumine (alpha-HSA/HSA) immunoreaction demonstrates the feasibility to use the microfluidic sensing system for immunosensing applications. In this case, a better discrimination between the bulk refractive index change and the layer formation can be made, resulting into higher accuracy and offering the prospect of being able to use the kinetics of the immunoreaction. The microfluidic sensing system shows an average phase resolution of 7 x 10(-5) x 2pi for different pairs of channels, which at the given interaction length of 4 mm corresponds to a refractive index resolution of 6 x 10(-8), being equivalent to a protein mass coverage resolution of 20 fg/mm(2). (C) 2004 Elsevier B.V All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据