4.7 Article

Improvement of direct determination of Cu and Mn in seawater by GFAAS and total elimination of the saline matrix with the use of hydrofluoric acid

期刊

TALANTA
卷 65, 期 2, 页码 402-407

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.talanta.2004.06.041

关键词

graphite furnace; seawater; Cu, Mu, HF precipitation of Ca and Mg

向作者/读者索取更多资源

Hydrofluoric acid, added to seawater, can assist in the removal of chloride in the drying step by precipitating fluoride salts, thus suppressing the chloride interference effects induced on the atomization signals of Cu and Mn. By adding HF to seawater before the analysis, MgF2 and CaF2 are precipitated at the bottom of the sampling flask, without precipitating Cu and Mn, and are consequently not introduced into the graphite furnace. Because sodium salts are eliminated at the pretreatment step. the whole seawater matrix is eliminated before the atomization of Cu or Mn. Therefore, the analyzed volume of seawater call be increased by using the multi-injection procedure without degradation of the limit of detection and risks of spectral interferences. The limit of detection obtained for Cu and Mn Lire 0.05 and 0.01 mug L-1, respectively, for a 50 muL analyzed seawater volume. (C) 2004 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据