PMMA; electrosynthesis; XPS; titanium; ion release; biomaterial
The performance of polyacrylic coatings as barrier films against corrosion of titanium-based orthopaedic implants was investigated. In particular, poly(methyl methacrylate) (PMMA) was electrosynthesized on titanium substrates by electro-reductive processes from aqueous monomer solutions. The obtained PMMA coatings were characterized by Fourier transform infrared spectroscopy (FTIR) and X-ray photoelectron spectroscopy (XPS). The effect of an annealing treatment on the morphology of coatings with respect to uniformity and porosity of films was assessed by scanning electron microscopy (SEM). An inductively coupled plasma-mass spectrometry (ICP-MS) technique was used for ion concentration measurements in ion release tests performed on TiAlV sheets modified with PMMA coatings (annealed and unannealed). Results indicated that the annealing process produces coatings with considerable anticorrosion performances.