The corner frequencies of the ECG amplifier for heart rate variability analysis

期刊

PHYSIOLOGICAL MEASUREMENT
卷 26, 期 1, 页码 39-47

出版社

IOP PUBLISHING LTD
DOI: 10.1088/0967-3334/26/1/004

关键词

cutoff frequency; electrocardiography; heart rate variability; high pass filter; low pass filter

向作者/读者索取更多资源

Heart rate variability (HRV) analysis is considered a popular method both in clinical and research fields. However, several ignored technical artifacts may falsify its measurements. The current study investigates the effects of corner frequencies of the ECG amplifier on the precision of RR-interval detection. Clear or noise-corrupted ECG records with predefined parameters consisting of 21 cycles were generated, played back through analog filters and data-logged on a Pentium-based computer with a DaqBoard2000 data acquisition card. 0.1-10 Hz second-order high pass and 20-100 Hz fourth-order low pass Bessel and Butterworth filters were used. The RR-intervals were measured between seven reference points of the ventricular complexes before and after filtering. High and low pass at every frequency cutoff and with both filter types results in correct RR-intervals within 1 ms error. However, a lower cutoff below I Hz is needed to maintain ECG morphology. AC interference or Gaussian random noise can falsify the measured RR-intervals up to 16 or 34 ms, respectively. These errors may be reduced to 1-4 ms with appropriate low pass filters. A frequency range of 0.5-20 Hz for the ECG amplifier can be sufficient for HRV analysis reducing the errors from AC interference or random noise.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据