4.4 Article

Ultra-sensitive carbon monoxide detection by using EC-QCL based quartz-enhanced photoacoustic spectroscopy

期刊

APPLIED PHYSICS B-LASERS AND OPTICS
卷 107, 期 2, 页码 275-283

出版社

SPRINGER
DOI: 10.1007/s00340-012-4949-1

关键词

-

资金

  1. NASA SBIR [NNX11CF34P]
  2. National Science Foundation (NSF) Engineering Research Center
  3. Welch Foundation [C-0586]
  4. Direct For Mathematical & Physical Scien
  5. Division Of Chemistry [1124677] Funding Source: National Science Foundation

向作者/读者索取更多资源

A quartz-enhanced photoacoustic spectroscopy (QEPAS) based sensor for carbon monoxide detection at ppbv levels was developed with a 4.65 mu m external-cavity quantum cascade laser operating both in continuous wave (cw) and pulsed modes. A 23-fold enhancement of the measured CO signal amplitude was obtained when water vapor, acting as a catalyst for vibrational energy transfer, was added to the targeted analyte mixture. In the cw mode, a noise-equivalent sensitivity (NES, 1) of 2 ppbv was achieved at a gas pressure of 100 Torr, for 1-s lock-in amplifier (LIA) time constant (TC), which corresponds to a normalized noise equivalent absorption coefficient (NNEA) of . In the pulsed mode, the determined NES and NNEA were 46 ppbv and , respectively, for a 3-ms LIA TC at atmospheric pressure with a laser scan rate of 18 cm(-1)/s and a 50 % duty cycle. An intercomparison between cw and pulsed QEPAS-based CO detection is also reported.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据